● 55*32mm jaw size for larger ground rods
● Autoranging ground resistance measurements from 0.01Ohm to 1000Ohm with 0.01Ohm resolution
● AC Leakage current range of 10mA to 30.00A with 50μA resolution
● Configurable alarm value for resistance and current to support your measurement
● Ergonomic mold, full rubber coating for slip proof
● Simplifies ground resistance measurements on multiple point ground systems
● Verification resistor confirm accuracy of meter
● Datalogging with 99 data points
● Complete with heavy duty case and verification resistor
- Earth Resistance:
- 0.01Ω ~ 0.99Ω 0.01Ω ±1%+0.01Ω
- 1.0Ω ~ 49.9Ω 0.1Ω ±1.5%+0.1Ω
- 50.0Ω ~ 99.5Ω 0.5Ω ±2%+0.5Ω
- 100Ω ~ 199Ω 1Ω ±3%+1Ω
- 200Ω ~ 395Ω 5Ω ±6%+5Ω
- 400Ω ~ 590Ω 10Ω ±10%+10Ω
- 600Ω ~ 1000Ω 20Ω ±20%+20Ω
- Leakage current:
- 0 ~ 80mA 50µA ±2.5%+1mA
- 80mA ~ 650mA 0.5mA ±2.5%+2mA
- 650mA ~ 4A 5mA ±2.5%+10mA
- 4A ~ 30A 10mA ±2.5%+20mA
- Min. Resolotion:50μA
- Conductor size:32 * 55mm
- Working Temperature:-10°C ~ 55°C
- Working Humidity:10%~80%
- LCD Size:28.5 x 47 mm
- Power Source:AA1.5Vx4
- Dimension:285 x 90 x 66 mm
- Weight:4kgs
|